Cold Sterilization

Cold sterilization applies to heat-sensitive items, such as non-metal, non-disposable vaginal speculums, anoscopes, scopes with light bulbs, such as Cystoscopes, and stainless-steel instruments.

1.  Use only activated high-level disinfectant at temperature 20° C or higher, and soak all items for a minimum of 10 hours.

2.  Keep disinfectant solutions covered and in a well-ventilated area.

3.  Thoroughly clean items with soap and water, rinse, and dry. Open and/or unlock all jointed items. Disassemble all items designed for disassembly.

4.  Immerse items completely in an activated, high-level, disinfectant solution, such as Cidex Plus, that is at a temperature of 20° C or higher, and soak for a minimum of 10 hours.

5.  Use sterile forceps to remove the items from the solution.

6.  Rinse items with sterile water, and place on a clean drape or towel.

7.  Use only sterilization disinfectant solutions. Keep solutions covered, and use in a well-ventilated area. Do not use solution after the expiration date determined according to manufacturer’s written recommendations marked on the container.